ETSI EN 303 203-1 covers the minimum characteristics of Medical Body Area Network System (MBANS),including the spectrum monitoring and access requirements, considered necessary in order to make the best use of theavailable spectrum within the 2 483,5 MHz to2 500 MHz frequency range and to avoid harmful interference betweenMBANS and other users of this band. It does not necessarily include all the characteristics which may be required by auser, nor does it necessarily represent the optimum performance achievable.
The types of devices that can belong to MBANSs are on-body and off-body medical sensors, patient monitoring devicesand medical actuators covered by the Medical Device Directive (Directive 93/42/EEC [i.7]).
The present document applies to the following MBANS applications which are considered to operate indoor:
- MBANS operating in the healthcare facility
- MBANS operating in the patient’s home
The present document contains the following basic technical characteristics of MBANS radio equipment which are also addressed in annex 2 of CEPT/ERC/REC 70-03 [i.2]:
- Healthcare facility MBANS with 1 mW maximum e.i.r.p. and not more than 10 % duty cycle over a maximumemission bandwidth of 3 MHz.
- Patient’s home MBANS with 10 mW maximum e.i.r.p. and not more than 2 % duty cycle over a maximumemission bandwidth of 3 MHz.
- Edition:
- 1.1.1
- Published:
- 11/01/2014
- Number of Pages:
- 42
- File Size:
- 1 file , 240 KB
- Note:
- This product is unavailable in Russia, Ukraine, Belarus